Research Article

The Effects of Increasing Mass Unloader Knee Orthosis and Two Kinds of Silicon and Polyethylene Pad on Pistoning Movement during Walking

Abstract

Introduction: Knee orthosis is used in different cases and can be seen pistoning movement in all orthosis. The purpose of this study was to evaluate the effects of increasing mass knee brace and two kinds of silicon and polyethylene (PE) pad on pistoning movement during walking, on health subject.
Material and Methods: A total of 20 healthy volunteers, including 10 men and 10 women, with an average age of 24.65 years, participated in this study. Subjects used unloader knee brace with PE and silicon pad, with and without weights in four phases, and at each phase, they engaged in 10 minutes’ walk. Before and after each phase, using a video camera, the brace displacement on limb was measured with a marker. Finally, the pistoning movement of the orthosis on the knee was measured using the Kinovea software.
Results:
This study showed significant differences between the four tests (P < 0.050). The mean migration of the orthosis with silicon pads in both genders was lower, compared to the average migration of the or hosis in the three other cases (0.65 cm), and the amount of migration knee brace with silicon pad with weights, PE pad, PE pad with weights was 4.1 ± 1.6, 3.1 ± 1.4, 6.3 ± 1.8 cm, respectively. An inverse relationship existed between the level of comfort and orthosis migration. Orthosis with silicon pad demonstrated the highest level of comfort and orthosis with PE pad, with weight showing the lowest level.
Conclusion: The results showed that the use of silicone pads inside the brace construction, as well as lighter than knee brace, reduced migration, and increased comfort level. Moreover, compliance to knee braces was improved.

Faso DR, Montgomer JB. The dynamic rotating functional knee orthosis concept. Orthotics Prosthet 1987; 41(3): 32-6.

Koo S, Andriacchi TP. The knee joint center of rotation is predominantly on the lateral side during normal walking. J Biomech 2008; 41(6): 1269-73.

Andriacchi TP, Mundermann A, Smith RL, Alexander EJ, Dyrby CO, Koo S. A framework for the in vivo pathomechanics of osteoarthritis at the knee. Ann Biomed Eng 2004; 32(3): 447-57.

Andriacchi TP, Briant PL, Bevill SL, Koo S. Rotational changes at the knee after ACL injury cause cartilage thinning. Clin Orthop Relat Res 2006; 442: 39-44.

Stergiou N, Ristanis S, Moraiti C, Georgoulis AD. Tibial rotation in anterior cruciate ligament (ACL)-deficient and ACL-reconstructed knees: a theoretical proposition for the development of osteoarthritis. Sports Med 2007; 37(7): 601-13.

Van de Velde SK, Gill TJ, Li G. Evaluation of kinematics of anterior cruciate ligament-deficient knees with use of advanced imaging techniques, three-dimensional modeling techniques, and robotics. J Bone Joint Surg Am 2009; 91(Suppl 1): 108-14.

Walker PS, Rovick JS, Robertson DD. The effects of knee brace hinge design and placement on joint mechanics. J Biomech 1988; 21(11): 965-74.

Regalbuto MA, Rovick JS, Walker PS. The forces in a knee brace as a function of hinge design and placement. Am J Sports Med 1989; 17(4): 535-43.

Lewis JL, Lew WD, Patrnchak CM, Shybut GT. A new concept in orthotics joint design-the Northwestern University knee orthosis system. Orthotics Prosthet 1983; 37(4): 15-23.

Singer JC, Lamontagne M. The effect of functional knee brace design and hinge misalignment on lower limb joint mechanics. Clin Biomech (Bristol, Avon) 2008; 23(1): 52-9.

Lew WD, Patrnchak CM, Lewis JL, Schmidt J. A Comparison of Pistoning Forces in Orthotic Knee Joints. Orthotics Prosthet 1982; 36(2): 85-95.

Brownstein B. Migration and Design Characteristics of Functional Knee Braces. J Sport Rehabil 1998; 7(1): 33-43.

France EP, Cawley PW, Paulos LE. Choosing functional knee braces. Clin Sports Med 1990; 9(4): 743-50.

Greene DL, Hamson KR, Bay RC, Bryce CD. Effects of protective knee bracing on speed and agility. Am J Sports Med 2000; 28(4): 453-9.

Miller JP, Vailas JC, Croce RV, Confessore R, Catlaw K. Dynamic Analysis of Custom-Fitted Functional Knee Braces: EMG and Brace Migration during Physical Activity. J Sport Rehabil 1999; 8(2): 109-22.

Van Leerdam N. The genux, a new knee brace with an innovative nonslip system. Chicago, IL: Academy of American Orthotists and Prosthetists; 2006.

Lamontagne M, Singer J, Xu L. Knee brace migration: determining the relative kinematics of the leg and brace during cycling. Proceedings of the 12th Annual Meeting of European Society of Movement Analysis for Adults and Children; 2003 Sep 11-13; Marseille, France. 2016.

Wojtys EM, Huston LJ. "Custom-fit" versus "offthe- shelf" ACL functional braces. Am J Knee Surg 2001; 14(3): 157-62.

Rast PD. A three-dimensional kinematic comparison of functional knee brace migration using three offthe- shelf designs [Thesis]. Pok Fu Lam, Hong Kong: The University of Hong Kong; 2000.

Liu SH, Mirzayan R. Current review. Functional knee bracing. Clin Orthop Relat Res 1995; (317): 273-81.

Vailas JC, Pink M. Biomechanical effects of functional knee bracing. Practical implications. Sports Med 1993; 15(3): 210-8.

Pierrat B, Millot C, Molimard J, Navarro L, Calmels P, Edouard P, et al. Characterisation of Knee Brace Migration and Associated Skin Deformation During Flexion by Full-Field Measurements. Exp Mech 2015; 55(2): 349-60.

Balsalobre-Fernandez C, Tejero-Gonzalez CM, del Campo-Vecino J, Bavaresco N. The concurrent validity and reliability of a low-cost, high-speed camera-based method for measuring the flight time of vertical jumps. J Strength Cond Res 2014; 28(2): 528-33.

Ogueta-Alday A, Morante JC, Rodriguez-Marroyo JA, Garcia-Lopez J. Validation of a new method to measure contact and flight times during treadmill running. J Strength Cond Res 2013; 27(5): 1455-62.

Files
IssueVol 10 No 2 (2016) QRcode
SectionResearch Article(s)
Keywords
Silicon pad Polyethylene pad Pistoning movement Knee brace

Rights and permissions
Creative Commons License This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.
How to Cite
1.
Khazaie E, Saeedi H, Vahab -Kashani R. The Effects of Increasing Mass Unloader Knee Orthosis and Two Kinds of Silicon and Polyethylene Pad on Pistoning Movement during Walking. jmr. 2017;10(2):48-52.